JWI-530 Waterproof Indicator Setting Instructions
Instructions for setting maximum weighing capacity and divisions
1.Press and hold the
key and
key to turn on the indicator and enter the capacity setting. First set the password "in.0000", press
key to accumulate the value. Press
to shift, enter "in.8529", then press
, the window will show "150.00 kg".
2.Press
key to move the decimal point, press
key to select the maximum range, and press
key to select the unit kg, g, lb, lb.oz/斤, PCS.
Method 2: If there is no desired capacity, press and hold
key to enter the free capacity setting. The leftmost digit of the window flashes. Press
to Shift right, press
key to modify the value; Press
key to select the unit kg, g, lb, lb.oz/斤, PCS. Press
to save and go to the division setting, otherwise press and hold
to cancel the current setting and go back to the weighing state.
3.When the model setting is completed and the division setting is entered, the window displays " 0.01kg".
4. Method 1: Press
key to select the common division, and press
key to move the decimal point to the right.
Method 2: If no division is required, press and hold
key to enter the free division setting (set according to the principle of 1/2/5 minimum scale).
The number on the far right of the window flashes. Press
key to shift to the right and press
key to modify the value; Press
key to enter the calibration state after the modification is completed, otherwise a long press will cancel the current setting and return to the capacity setting.
5. When the calibration is completed after the division setting, the window shows "CAL". Press
key to enter the calibration, and long press
key will exit the calibration and return to the weighing state.
General Parameter Setting
1. Ingrese los parámetros: mantenga presionado
durante el estado de pesaje o presione
para ingresar la configuración cuando se encienda
2. Seleccione el parámetro: presione
la tecla para recorrer los parámetros después de ingresar a la pantalla de parámetros
3. Establecer parámetros: después de seleccionar el parámetro, presione
para ingresar a la interfaz de configuración de parámetros, presione
para seleccionar el contenido del parámetro
4. Guardar configuración: después de configurar, presione
para guardar la configuración y salir de la pantalla de configuración. Luego presione
para volver al estado de pesaje